BANTE900P PORTABLE MULTIPARAMETER WATER QUALITY METER
14
MULTI-POINT CALIBRATION:
2.1 Make sure that you have selected 2 to 5 points calibration in the setup menu.
2.2 Repeat steps 1.2 to 1.5 above. When the first calibration point is completed, the display shows
“CAL2”. The meter prompts you to
continue with second point calibration.
2.3 Rinse the pH electrode with distilled water.
2.4 Dip the electrode into the next standard buffer solution (e.g. pH4.01). The meter automatically sense the current calibration
solution and begin calibrating,
“Calibration” indicator
begins flashing.
2.5 Wait for the measured value to stabilize, the display shows
“CAL3”. The
meter prompts you to continue with third point calibration.
2.6 Repeat steps 2.3 to 2.4
above until the display shows “END”. The meter
automatically returns to measurement mode. Calibration is
completed.