Adjust (continued)
Displayed values (example)
At 0…10 V
At 4...20 mA
Set end point
►
Bring float to the new end point.
9.89 V
19.13 mA
►
Press
b
for at least 2 s.
> 2 s
b
⇒
The new end point is set with the last valid
end value.
10.00 V
20.00 mA
Adjust end value
►
The end value can be changed using
a
and
b
. The
gradient of the curve changes
10.00 V
20.00 mA
8.00 V
16.80 mA
►
End calibration procedure: Briefly press
a
and
b
simultaneously (< 1 s).
⇒
Set position value is saved.
< 1 s
a
b
2.00 V
6.00 mA
8.00 V
16.80 mA
End adjustment
►
Briefly press
a
and
b
simultaneously (< 1 s).
< 1 s
a
b
10.50 V
3.60 mA
⇒
The current position value is displayed once the
buttons are released.
7.63 V
18.56 mA
Any of the individual steps for settings can be
selected. The adjustment process can be
ended at any time.
a
a
b
b
BTL7-A/C/E5 _ _ -M _ _ _ _ -SF-F _ _
Magnetostrictive Linear Position Sensor – Rod Style
Summary of Contents for BTL7-A5-M Series
Page 2: ...www balluff com...
Page 3: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Betriebsanleitung deutsch...
Page 4: ...www balluff com...
Page 26: ......
Page 27: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ User s Guide english...
Page 28: ...www balluff com...
Page 50: ......
Page 51: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Notice d utilisation fran ais...
Page 52: ...www balluff com...
Page 74: ......
Page 75: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Manuale d uso italiano...
Page 76: ...www balluff com...
Page 98: ......
Page 99: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Manual de instrucciones espa ol...
Page 100: ...www balluff com...
Page 122: ......
Page 123: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Instrukcja obs ugi polski...
Page 124: ...www balluff com...