english
14
6
Calibration procedure (continued)
6.3.2 Adjust
The detailed procedure for adjusting is
The factory-set null point and end point are replaced by a
new start point and end point and the associated output
values can be adjusted. The start and end values can be
adjusted as you like up to the limits.
Steps:
►
Move float to the new start position.
►
Read new start point by activating the programming
inputs.
►
Set the desired start value by activating the
programming inputs.
New null point
Before
After
New start
value
Fig. 6-4: Adjust a new start value
►
Move float to the new end position.
►
Read new end point by activating the programming
inputs.
►
Set the desired end value by activating the
programming inputs.
New end point
Before
After
New end value
Fig. 6-5: Adjust a new end value
6.3.3 Reset
Restoring the BTL to its factory settings.
The detailed procedure for the reset is
BTL7-A/C/E5 _ _ -M _ _ _ _ -SF-F _ _
Magnetostrictive Linear Position Sensor – Rod Style
Summary of Contents for BTL7-A5-M Series
Page 2: ...www balluff com...
Page 3: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Betriebsanleitung deutsch...
Page 4: ...www balluff com...
Page 26: ......
Page 27: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ User s Guide english...
Page 28: ...www balluff com...
Page 50: ......
Page 51: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Notice d utilisation fran ais...
Page 52: ...www balluff com...
Page 74: ......
Page 75: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Manuale d uso italiano...
Page 76: ...www balluff com...
Page 98: ......
Page 99: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Manual de instrucciones espa ol...
Page 100: ...www balluff com...
Page 122: ......
Page 123: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Instrukcja obs ugi polski...
Page 124: ...www balluff com...