www.balluff.com
13
english
6
Calibration procedure (continued)
6.3
Calibration procedure overview
6.3.1 Teach-in
The factory-set null point and end point are replaced by a
new null point and end point. The null point and end point
can be set separately, the output gradient changes.
The detailed procedure for teach-in is described
Steps:
►
Move float to the new null position.
►
Read new null point by activating the programming
inputs.
⇒
The current end point remains the same.
New null point
Before
After
Fig. 6-2: Reading new null point
►
Move float to the new end position.
►
Read new end point by activating the programming
inputs.
⇒
The current null point remains the same.
New end point
Before
After
Fig. 6-3: Reading new end point
Only with BTL7-C/E...:
►
The gradient of the current output can be inverted by
activating the programming inputs.
⇒
For example, a rising output gradient is changed to
a falling gradient. The voltage outputs are not
inverted.
BTL7-A/C/E5 _ _ -M _ _ _ _ -SF-F _ _
Magnetostrictive Linear Position Sensor – Rod Style
Summary of Contents for BTL7-A5-M Series
Page 2: ...www balluff com...
Page 3: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Betriebsanleitung deutsch...
Page 4: ...www balluff com...
Page 26: ......
Page 27: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ User s Guide english...
Page 28: ...www balluff com...
Page 50: ......
Page 51: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Notice d utilisation fran ais...
Page 52: ...www balluff com...
Page 74: ......
Page 75: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Manuale d uso italiano...
Page 76: ...www balluff com...
Page 98: ......
Page 99: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Manual de instrucciones espa ol...
Page 100: ...www balluff com...
Page 122: ......
Page 123: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Instrukcja obs ugi polski...
Page 124: ...www balluff com...