CL 5
3-23
The following instrument settings are automatically
adjusted when the Setup file is activated:
View Mode
Gain
Measured Value display resolution
Minimum Alarm Setting
Maximum Alarm Setting
Zero Offset – for contact probes only
Velocity
Nominal Thickness
Note:
While Custom Setups can be created in any
instrument, the parameter settings, which may be
modified and then saved in a Custom Setup, depend
on the instrument’s configuration. In a base model
instrument, only the setup name, material velocity
value (determined via calibration), alarm thickness
values, and UT settings related to the probe setup
can be modified and stored in a custom setup.
Setting Up the CL 5
Configuring the Instrument‘s Display