Info/Help
Error Counts
[0]
U
Description
This menu provides a diagnostic view of meter performance . Below are
several system diagnostic counters and their definitions . Use discretion when
interpreting these counters . These values could be altered during system setup
or when using the verification device . We suggest that you reset these counters
before you start monitoring your system and look for conditions possibly
affecting performance .
Sensor
The number of times an invalid sensor condition has been observed .
Empty Pipe
The number of times an empty pipe condition has been observed by the meter .
Full Scale
The number of times the flow has exceeded the Full Scale Flow setting .
Totalizer
The number of times the totalizers have exceeded limits of the meter .
Pulse Sync.
The number of times the pulse outputs have fallen out of synchronization .
ADC Interrupt
The number of times an analog input measurement has been missed .
ADC Range
The number of times the analog input measurement range has been exceeded .
System Error
A diagnostic system message indicating the reason for a system reset .
System Resets
The number of times the meter has been reset .
System Reset
ID
Diagnostic information about a system reset as a result of expired
internal timers .
Token Errors
Indicates the number of parameter copies from a memory token that failed to
be copied to the meter .
Checksum
For MID meters . If data corruption to the meter memory occurs, this “counter” indicates the
memory region that is corrupted .
PowerUp Counter
[Not applicable]
U
The number of times that the unit has been powered on .
Power Off Totalizer
[Not applicable]
U
The length of time that the unit has been without power .
Version Info
[Not applicable]
U
The current software version .
Serial Number
[Not applicable]
U
The manufacturing serial number in the format YYMM####, where YYMM indicates year and
month of manufacturing and #### indicates the sequence number .
Meter Tag Name
For Profibus – This parameter is only programmable over external Profibus communications .
Daughterboard
Information
Describes current version of attached daughterboard .
OTE:
N
RS485 daughterboard is not recognized because it is a pass-through device rather than
an intelligent protocol converter like HART-to-Modbus .
Polarization Voltage
Diagnostic voltage to help determine if the meter or application is performing optimally .
Using the Main Menu Programming Options
Page 55
March 2022
MAG-UM-01272-EN-21