Advanced
Empty Pipe Cal.
[Default]
S
Fluid conductivity impacts the performance of empty pipe measurements . If you require empty pipe
detection, you should perform this empty pipe calibration procedure .
Before starting the empty pipe calibration, verify that empty pipe detection is enabled . Also, run both
the empty pipe and the full pipe calibration procedures .
Calibrating an Empty Pipe
Before calibrating an empty pipe, verify that the pipe is empty .
To calibrate with an empty pipe, follow these steps from the
Advanced
menu:
1 . Select
Empty Pipe Cal
to view the
Calibration
menus .
2 . Select
Cal. Empty Pipe
to view the
Empty Pipe Calibrate
menu .
3 . To enable calibration, place the cursor on the calibration enable line and press [
E
] .
4 . Wait 30 seconds for voltage measurement to stabilize .
5 . To save the setting, place the cursor on
Exit with Save
and press [
E
] .
Calibrating a Full Pipe
Before calibrating a full pipe, verify that the pipe is full .
To calibrate with a full pipe, follow these steps from the
Advanced
menu:
1 . Select
Empty Pipe Cal
to view the
Calibration
menus .
2 . Select
Cal. Full Pipe
to view the
Full Pipe Calibrate
menu .
3 . Enable calibration by placing the cursor on the calibration enable line and press [
E
] .
4 . Wait 30 seconds for voltage measurement to stabilize .
5 . To save the setting, place the cursor on
Exit with Save
and press [
E
] .