Error description
Possible causes
Remedy
Restat the scanner, restart
B9
Scan
application and run an-
other calibration procedure
•
Calibration plate damaged or
scratched
•
Replace the calibration plate
•
Wrong spacer used
•
Use long spacer
•
If none of the above solves the
problem, contact E.G.S. srl cus-
tomer service
14
The
B9 Scan
applications
runs correctly but the pro-
jector's light does not turn
off after a calibration
•
Projector failure
•
Unplug the device immediately
and contact E.G.S. srl customer
service