Azbil Corporation
Operation using the data setting device
Model MCB10A 5-7
5-3 : Description of MEASURING MODE
5-3-1 : Display overview
% flow rate display
10 0.0
%
7 . 0 6 9
8 8 8 8 8 8 8 8
m3/h
1st line (Main display): 7-segment 4-digit display
% flow rate (%)
2nd line: Actual flow rate display (Significant
value of 5 digits)
3rd line: Totalized value display (Significant
figure of 8 digits)
Actual flow rate display
7.06 9
%
RATE
1 0 0 . 0
8 8 8 8 8 8 8 8
m3/h
1st line (Main display): 7-segment 4-digit display
Actual flow rate (RATE)
2nd line: % flow rate display (Significant figure
of 4 digits), unit of actual flow rate
3rd line: Totalized value display (Significant
figure of 8 digits)
Totalizer display
1 2 3 4 5 6 7 8
7 . 0 6 9
m3/h
1 0 0 . 0 %
TOTAL
1st line (Main display): 7-segment 8-digit display
Totalized value (TOTAL)
2nd line: Actual flow rate display (Significant
figure of 4 digits)
3rd line: % flow rate display (Significant figure
of 4 digits)
Totalization is not performed, when the output selection is set to the contact output.
However, the previous value is displayed as the totalized value.
* Details on display
% flow rate display:
The % flow rate display range is from -115.0% to 115.0%.
A value up to the first decimal place is displayed. The
position of the decimal point is fixed.
The integer part to be displayed has up to three digits (0 to
115).
In the main display, unnecessary zeros are deleted (but are
not deleted in the sub display).
Example) 019.8%
g
19.8%
-000.5%
g
-0.5%
The position of the negative sign (-) is fixed. (The positive
sign is not displayed.)
Actual flow rate display: The flow rate displayed in the actual flow rate display is up
to 115% of the range or the equivalent.
However, if the flow rate equivalent to 115% of the range
exceeds the range of significant figure, the highest value
(e.g., 9.999) will be displayed.
In the main display, unnecessary zeros are deleted (but are
not deleted in the sub display).