Operation using the data setting device
Model MCB10A 5-29
5-6-5 :
Setting and changing compensation coefficient
This function is used to set or change the compensation coefficient which is used to
multiply the output flow rate as required.
Set range:
0.10000 to 5.9999
Default:
1.0000
Set and change a compensation coefficient in accordance with the following
procedure:
Step
Procedure
Screen
1
Enter ENGINEERING MODE in accordance
with the entry into ENGINEERING MODE
(see section 5-4-10). Then press the
h
or
i
key to display the screen at right.
12.3 %
# COEFFICIENT
1.0000
2
Press the
g
key to move the cursor to the
position under a desired value to be set or
changed.
12.3 %
# COEFFICIENT
1.0000
3
Using the
h
or
i
key, change the value to the
desired one to be set.
12.3 %
# COEFFICIENT
1.0050
4
Press the
g
key to move the cursor to the
position under #.
Press the MODE key to return to the
MEASURING MDOE and to save data.
12.3 %
# COEFFICIENT
1.0050
CAUTION
You have only ten minutes to return to MEASURING MODE to save the new
value before the system resets it to the previously saved value.