Operation using the data setting device
Azbil Corporation
5-6
Model MCB10A
This section describes keys on the data setting device.
Name
Description
MODE key
•
Enters OPERATOR'S MODE.
•
When parameters and configured data have
been changed in ENGINEERING MODE or
MAINTENANCE MODE, press this key to save the
data.
Right shift key
•
Moves the cursor to the right.
Decrement key
• Changes the parameter at a cursor position.
•
Displays the previous screen.
* OPERATOR'S
MODE
Cursor
If the key is pressed, when
the cursor is placed at the
upper left end (*, #, >),
the screen will change.
* DAMPING
001.0 S
Cursor
If the key is pressed, when
the cursor is placed at a
number, the number is
decremented.
# 1.0000 m/s
SPAN 07. 069 m
3
/h
Cursor
If the cursor is placed at a
decimal point, the decimal
point moves rightward.
Increment key
• Changes the parameter at a cursor position.
•
Displays the following screen.
* OPERATOR'S
MODE
Cursor
If the key is pressed, when
the cursor is placed at the
upper left end (*, #, >),
the screen will change.
* DAMPING
001.0 S
Cursor
If the key is pressed, when
the cursor is placed at a
number, the number is
incremented.
# 1.0000 m/s
SPAN 07. 069 m
3
/h
Cursor
If the key is pressed, when
the cursor is placed at a
decimal point, the decimal
point moves to the left.
* AUTO ZERO
READY
Cursor
If the cursor is placed over
READY, pressing the key
starts operation.