Operation using the data setting device
Model MCB10A 5-19
5-5-7 :
Selecting flow rate to be displayed in the main display
Select the flow rate to be always shown in the main display. The flow rates other than
that selected for the main display are shown in the sub displays. Thereby, three flow
rates can always be monitored.
Set
Description
%
% flow rate
RATE
Actual flow rate
TOTAL
Totalized value
% (% flow rate): Displays % flow rate
RATE (Actual flow rate)
Setting range:
%, RATE, TOTAL
Default:
%
Select the flow rate to be shown in the main display in accordance with the following
procedure:
Step
Procedure
Screen
1
The screen at right is a screen display example
in MEASURING MODE (measurement state).
Press the MODE key.
20.0 %
01.94 m
3
/h
00069401
2
The OPERATOR'S MODE screen appears
for approx. two seconds and then the damping
setting screen appears.
20.0 %
* OPERATOR'S
MODE
20.0 %
* DAMPING
005.0 s
3
Press the
h
or
i
key to display the screen as
shown on the right.
20.0 %
* DISP SELECT
%
4
Press the
g
key to move the cursor to the
position for the flow rate display method (%,
RATE, TOTAL). The screen at right shows
an example where % flow rate has been set for
the main display.
20.0 %
* OPERATOR'S
MODE
5
Press the
h
or
i
key to select a flow rate
display to be set. The screen at right shows
an example where RATE (actual flow rate)
display has been selected.
0.30
* DISP SELECT
RATE
RATE
6
Press the
g
key to move the cursor to the
position under *.
Press the MODE key to return to the
MEASURING MODE and to change to the
set flow rate display.
0.30
* DISP SELECT
RATE
RATE
CAUTION
You have only ten minutes to return to MEASURING MODE to save the new
value before the system resets it to the previously saved value.