Operation using the data setting device
Azbil Corporation
5-48
Model MCB10A
5-7-5 :
Configuration of CALIBRATION MODE
Introduction
CALIBRATION MODE has the following setting and adjustment items:
Configuration of CALIBRATION MODE requires a dedicated calibrator.
Wrong operation may hinder accurate measurements of the flow rate. To operate in
this mode, contact an Azbil Corp. representative.
Item
Content
Screen
CAL I.OUT
Adjusts 4 mA analog
current output.
12.3 %
> CAL I.OUT OFF
LOW
4.000 mA
LOW
4.000 mA
CAL I.OUT
Adjusts 20 mA analog
current output.
12.3 %
> CAL I.OUT OFF
HIGH
20.000 mA
HIGH
20.00 mA
CAL P.OUT
Adjusts 90 Hz pulse
output.
12.3 %
> CAL P.OUT OFF
FREQ
90 Hz
FREQ
90 Hz
CAL GAIN
Adjusts 0 m/s gain.
12.3 %
> CAL GAIN OFF
ZERO
READY
ZERO
CAL GAIN
Adjusts 5 m/s gain.
12.3 %
> CAL GAIN OFF
5.0 m/s
READY
5 m/s