4 MAINTENANCE AND TROUBLE SHOOTING
4.1 Maintenance
The Axetris LGD F200P2 gas sensor modules will not require routine maintenance when the installation
and operation instructions described in this document are carefully applied. The surface of the instrument
can be cleaned with a soft tissue and the use of isopropyl alcohol. Any gas filtration and conditioning
equipment must be checked and replaced periodically according to manufacturer specifications.
If the measurement cell gets contaminated or damaged, please contact Axetris for cell maintenance and
replacements. As technical training is necessary for this procedure, we recommend setting up a system
exchange program. Please contact Axetris for details.
In case of problems, recheck all the recommendations presented in this document and be sure they have
been followed and fully applied. If the observed problems persist, contact your Axetris sales person.
It is important that this device is serviced only by adequately trained and
qualified personnel.
This instrument contains electronic components that might be damaged by static
electricity. The device cover should never be removed as it might damage the optics
and/or electronics.
Axetris gas sensor modules contain metal and elastomeric sealing parts. The customer is
responsible for checking the compatibility of those materials with the gas applied to the
device. Not observing these recommendations might lead to personal injury or death.
It is recommended that the user checks the device regularly to ensure that it is leak-free.
Sealings may change their characteristics with age, exposure to process gas, temperature
and pressure.
If it is necessary to remove the device from the system, unplug the electrical power before
any other manipulation.
Before removing the device from the system, purge thoroughly with a dry inert gas such as
Nitrogen before disconnecting the gas fittings. Ensure that the piping is de-pressurized.
Failure to correctly purge and de-pressurize the device could result in fire, explosion, or
severe injury.
Summary of Contents for LGD F200P2
Page 1: ...Operating Integration Instructions Laser Gas Detection OEM Module LGD F200P2 Axetris AG...
Page 57: ...Axetris AG LGD F200P2 Operating Instructions Rev D 19 09 2018 57 60 Notes...
Page 58: ...Axetris AG LGD F200P2 Operating Instructions Rev D 19 09 2018 58 60 Notes...
Page 59: ...Axetris AG LGD F200P2 Operating Instructions Rev D 19 09 2018 59 60 Notes...
Page 60: ...Axetris AG LGD F200P2 Operating Instructions Rev D 19 09 2018 60 60 Notes...