Axetris AG
LGD F200P2 Operating Instructions - Rev D
19.09.2018
23/60
Status and error codes
Status and error information is shown in the lower section of the GUI window. In normal operation the
following status window is displayed:
If the sensor transmits a warning or an error, the color of the status indicator changes to yellow (warning)
or red (error). It shows a short description of the sensor status/error and the corresponding hexadecimal
error code [0x…]. The complete list of status, warning and error messages can be found in Annex A and
B.
Example:
The GUI window only shows the present status, warning or error message with the highest priority.
To see a complete list with the history
of all status and error information,
select
Sensor\error status
in the
menu. This will show a window with all
the codes transmitted from the sensor
since the interfacing session started.
For every code, a counter and time
stamp for the latest status/error
received is displayed. Status, warning
and error codes are detailed in Annex
A and B.
Figure 18
History list of status and error codes
Sensor parameters
The sensor offers the possibility of readjusting
measured values, and configuring and
adjusting the analog output (4 -20 mA or 0 -
5 V) or the integration time. Select
Sensor /
Change parameters
in the pull down menu to
open the parameter list, as displayed in
Figure 19
. Parameters that are being
changed appear in orange color until they are
saved by pressing the
Save to Sensor
button.
The values altered here are saved in the
firmware and used with the analog interface.
Please note that these parameters
are saved in the EEPROM of the instrument
that has a limited number of writing cycles
of 10,000.
Figure 19
Sensor user parameter settings
2.12.5.1 Calibration adjustment
Adjustment of the sensor calibration
Customer offset [ppm]
and
Customer span [ppm/ppm]) allows
correcting the measured
values with the linear function:
Summary of Contents for LGD F200P2
Page 1: ...Operating Integration Instructions Laser Gas Detection OEM Module LGD F200P2 Axetris AG...
Page 57: ...Axetris AG LGD F200P2 Operating Instructions Rev D 19 09 2018 57 60 Notes...
Page 58: ...Axetris AG LGD F200P2 Operating Instructions Rev D 19 09 2018 58 60 Notes...
Page 59: ...Axetris AG LGD F200P2 Operating Instructions Rev D 19 09 2018 59 60 Notes...
Page 60: ...Axetris AG LGD F200P2 Operating Instructions Rev D 19 09 2018 60 60 Notes...