The cathode should be on the right. The anode should be on the left.
Lower the green lever on the left. This will lower the nylon screw over the DUT area on
the right, and gently lock the DUT in place in the cavity:
(It is possible to use a screwdriver to adjust the positioning of the nylon screw. This will
vary the maximum mechanical pressure applied to the DUT.)
Once the DUT is installed in the socket, close the main lid. It must be closed for proper
operation.
AVX-CA-KMPB-TO39
The AVX-CA-KMPB-TO39 accepts diodes in three-pin TO-39 packages. The anode
and cathode must be opposite each other across the center of the package. The
anode-to-cathode pin spacing is 0.2” (or 5.08 mm). Pin sockets are provided for the
35
Summary of Contents for AVR-EB2A-B
Page 44: ...The installed DUT is shown below 44 ...
Page 46: ...The installed DUT is shown below 46 ...
Page 48: ...It is still below 300 ps at 10 mA 10 mA as well 48 ...
Page 56: ...PCB 158R4 LOW VOLTAGE POWER SUPPLY ...
Page 65: ...TEST JIG WIRING AVX CA AXPOST ...
Page 67: ...TEST JIG WIRING AVX CA SOD123 QTKA ...
Page 69: ...TEST JIG WIRING AVX CA SOT23B ...
Page 70: ...CUSTOMIZED DAUGHTERBOARD AVX CA SOT23B ...
Page 71: ...PERFORMANCE CHECK SHEET 71 ...