WITH NO DUT
If a piece of busbar wire is installed in the axial pocket, the basic speed of the test
socket may be judged. With I
F
= +100 mA and I
R
= -100 mA, a 10%-90% switching time
of 709 ps observed:
The inductance of the busbar and the socket degrades the transition time of the pulse
from the mainframe. If the mainframe OUT is fed directly into an oscilloscope
(bypassing the test jig), transition times of less than 200 ps are observed: