E1070 Service Manual
31
3.4 Service Menu
8.
Toggle between the choices by using the
TARE
or
UNITS
key and press the
ENTER
key to accept the choice…
C-
Unit
is displayed.
9.
Press the
SELECT
key…
UnitS
is displayed.
STABLE (Stability window)
Use this item to define the stability window in terms of ±X divisions for a period of time,
in seconds, you set.
1.
With
UnitS
displayed, press the
UNITS
key…
StAbLE
is displayed.
2.
Press the
key…
div.
is displayed.
3.
Press the
key…
The current division size is displayed. If a weight changes less than this
number of divisions in the time period you select in the next steps, the
motion light turns off and the weight is considered stable.
You choices are
0.25
,
0.5
,
1
,
3
and
CUSt.
(custom).
4.
Scroll through the choices by using the
TARE
or
UNITS
key and press the
ENTER
key to accept the displayed choice…
div.
is displayed for any choice other than
CUSt.
If you pick
CUSt.
go to
step
. If you picked any other division size, go to step
.
If your new custom unit is larger than one CAL UNIT, then you key in how many CAL
UNITS
make up 1 new custom unit. For example 1 TON = 2000 pounds so with
pounds selected as our CAL UNIT we would key in 2000 for the multiplier.
If your new custom unit is smaller than one CAL UNIT, then you divide one cal unit by
the number of custom units it takes to make up a single CAL UNIT. Multipliers are
limited to a total of seven digits by the display.
Example #1:
16 ounces = 1 pound.
Do the math: (one cal unit / number of custom units = the multiplier)
1/16=0.0625
So with pounds selected as our CAL UNIT we would key in 0.0625 for the multiplier.
Example #2: 1000 Grams = 1 KG.
Do the math: (one cal unit / number of custom units = the multiplier)
1/1000=0.001
So with KG selected as our CAL UNIT we would key in 0.001 for the multiplier.
Summary of Contents for E1070
Page 1: ...Model E1070 Indicator Service Manual NORTH AMERICA AWT35 500106 Issue AF March 2010 ...
Page 6: ...6 E1070 Service Manual ...
Page 51: ...E1070 Service Manual 51 3 4 Service Menu Table 3 1 Printalbe Characters Chart ...
Page 146: ...146 E1070 Service Manual 8 Network Connections ...
Page 150: ...150 E1070 Service Manual 10 2 Complete Service Menu ...
Page 153: ...E1070 Service Manual 153 11 3 Wall Mount Application ...
Page 154: ...154 E1070 Service Manual 11 4 System Block Diagram ...
Page 155: ...E1070 Service Manual 155 11 5 Keypad Schematic and Power Supply Bd P N 60071 0016 ...
Page 158: ...158 E1070 Service Manual 11 8 I O External Calbe Identification Pin outs ...
Page 162: ...162 E1070 Service Manual ...
Page 163: ......