User manual
– SYSTOE
14 168
Rev. G
Page 32/38
Control of zero pressure
Sensor pressure
Press.: 252mmHg
RAZ RET.
Press on RAZ
The value on the manometer should be 0 mmHg
Control of the 100 mmHg pressure
With the bladder of the sphygnomanometer, input a pressure equal to 100
mmHg on the manometer. Read the value displayed on the SysToe.
The SysToe passes the test when the displayed pressure on the SysToe is
equal to 100 mmHg
3 mmHg.
If the SysToe fails the tests, it should be calibrated by Atys.
14.4.7 Calibration
The following calibrations can be performed:
Calibration of the pressure sensor.
Calibration of the deflation of the occlusion cuff.
The calibration must be performed by Atys.
Fully loosen the screw
RAZ