User manual
– SYSTOE
14 168
Rev. G
Page 6/38
1.7
Electrical safety
To avoid the risk of electrical shock, do not remove the covers of the
charger.
To avoid the risk of electrical shock, contact a qualified service
representative for service.
Before use, control the sensor and its cable for visible damages. Never
use the sensor if cracks or other damages are visible.
Disconnection from the main supply: disconnect the 2 poles
simultaneously: unplug the charger.
Do not position device so that it is difficult to disconnect from the mains.
In case of breakdown of the SysToe, please contact your Atys dealer.
1.8
Environmental protection
Do not wasp the SysToe and its accessories. They can be partially
recovered and re-used.
The wasting of the package and the SysToe must be performed
according the national regulations.
2
Intended use
The
SysToe
is designed to measure the toe or the digit systolic pressure in
a non-invasively way in order to assess peripheral vascular disease. The
SysToe
must be used in hospitals, physician office or medical facilities.
3
Principle of operation
3.1
Introduction
The
SysToe
inflates the occlusion cuff up to the preset occlusion pressure.
Then, it deflates the occlusion cuff slowly until the pressure in the cuff
reaches 10 mmHg.
The pressure in the cuff and the PPG (Photo PlethysmoGraphy) signal are
recorded during deflation. The pressure at the time of resumption of arterial
inflow is the systolic pressure.
Resumption of arterial inflow is detected by the PPG sensor.