14.4.4 Control of the occlusion cuff leakage
Install the occlusion cuff on a tube of 10-20mm diameter.
Press on START
The device will inflate the occlusion cuff and test for leakage during 15s.
The final leakage value is displayed.
The result of the test is displayed as “test passed” or “test failed”.
Occlusion cuff
Initial Pres.:305mmHg
Final Pres.:290mmHg
Falling Pres.: 15mmHg
Test passed
Chrono: 0s/15s
START RET.
14.4.5 Control of the sensor cuff leakage
Install the sensor cuff on a tube of 10-20mm diameter as above.
Press on START
The device will inflate the sensor cuff and test for leakage during 15s.
The final leakage value is displayed.
The
result of the test is displayed as “test passed” or “test failed”.
Sensor Cuff
Initial Pres.:305mmHg
Final Pres.:290mmHg
Falling Pres.: 15mmHg
Test passed
Chrono: 0s/15s
START RET.
14.4.6 Control of the pressure sensor calibration
The control of the pressure sensor calibration can be performed by any
qualified technician.
It is advised to perform this control once a year.
The calibration of the pressure sensor must be performed by Atys.
Required equipment: sphygmomanometer that must be calibrated by an
accredited laboratory.
The battery of the SysToe must be charged. The following message should
not be displayed.
Connect a sphygnomanometer on the occlusion cuff output.