User manual
– SYSTOE
14 168
Rev. G
Page 33/38
15
Regulations
EC: Class IIa device
Canada: Class 2 device
16
Technical data
16.1 Electrical characteristics
Internally powered equipment with a NiMH battery.
Type: 18 522 Systoe battery pack 2500mAh
Replacement interval: 1 year
Charge time: 5h
Discharge time standby: 3h
Discharge time operating: 150 examination cycles.
16.2 Physical specifications
Dimensions: 15 cm x 10 cm, height: 5.5 cm
Weight: 0.5 kg
16.3 Pressure sensor
Measurement and display range: 0-330 mmHg
Accuracy: +/- 3%, above 70mmHg and +/- 2mmHg below 70mmHg
16.4 PPG Sensor
Infrared type sensor.