User manual
– SYSTOE
14 168
Rev. G
Page 30/38
14.4 Troubleshooting
In case of trouble and at least once a year,
Verify the unit integrity and the sensor/cuff connections
Operate the following software assisted tests from the diagnostic
module.
14.4.1 Main menu
From the acquisition screen, press on
to display the main menu.
Press on DIAG to display the diagnostic menu.
14.4.2 Control of the battery charge and temperature
Diagnostic Battery
VB 385 Tp 25 Ch 80%
DC adaptor
not connected
NOT CHARGING
RET.
VB: Current battery level.
VBatt > 450mV at end of charge
Tp: Internal device temperature in °C < 60°C
Ch: Actual charge level.
RET.: Return to diagnostic main menu
14.4.3 Control of the sensor (photo detector)
Press on INFO. Apply the sensor on a digit. Press on RESET.
The electrical signal value is displayed.
Diagnostic PPG
Sens.
signal = 550
300 < signal < 700
Test passed
RET.
The result of the test is displayed:
Test passed
Test failed
RET.: Return to diagnostic main menu
DIAG