93
In order to create a batch name and enter calibration information, perform the following steps:
a. Select Create from the Thermocouple Batch pull-down under the Calibration menu.
b. Click on New or Add and enter the Batch name. Be sure to use a unique name for each batch.
c. Enter the appropriate information.
d. Enter the Temperature and Offset Correction in the appropriate window.
e. Click Add.
f. Repeat these steps for each temperature point in the Correction Data window.
g. Click OK to complete the batch creation process.
Previously created batches can be edited, viewed, or deleted by choosing the appropriate option from
the Thermocouple Batch pull-down from the Calibrate menu.