7.6.3 Report/Graphing Options
The Report/Graphing Options screen allows the user to select a specific report or graph and configure it
to display test data in the most useful form. The screen also allows selected data to be exported into a
Common Separated Variable (.csv) format for spreadsheet use.
1. Select View from the main menu, then either Previous Test(s) or Running Test(s) to access the
Report/Graphing Options screen.
2. Under the Report(s) tab, check the boxes for the data to be shown in the report.
a. If the Short Form box is checked, only the following information will be included in the report:
Specimen Information, Test Condition, Test Results, Intermediate Creep Values (if appropriate),
Post Test Comments, Operator Identification.
b. If the Long Form box is selected, all the data listed below the Long Form box will automatically
be selected. If not all this information is desired, items can be selected individually.
NOTE: Any of the listed items can be selected, and they will be presented in tiled formatting. The system
will make inappropriate selections unavailable, such as a creep graph on a stress rupture test.