36
6.2.8 Miscellaneous Setup
There are miscellaneous items that must be specified. These are established at the time of installation,
but may need to be changed for various reasons. Contact ATS at +1-724-283-1212.
Frame Motion Delays:
Delays are sometimes needed between the control input and the resulting action
in order to protect the hardware. In most cases, the unload value is not used and the box is left clear.
The unload time at the end of the test is the amount of time that the drawhead will run to remove the
load at the end of the test. The time must be long enough to fully remove the load; however, the time
must not be excessive or damage can result to the tester, load train, extensometer, or furnace.
On frames with motorized drawheads or elevators, a manual control delay is essential. If the motor has
not stopped before the reverse motion command is given, the motor may continue to run in the original
direction. If both commands are given simultaneously, the motor may be drawing excessive current and
will be damaged. This can also cause protective devices to open. To avoid these issues, a three-second
delay is normally used. When a reverse direction is specified, the frame motion controls on the system
console will be inactive for three seconds, allowing the motor time to stop and reverse safely.
NOTE: A casual operator may be unaware that this delay is a safety measure rather than a sign of
equipment malfunction. It is essential for operators to understand this safety feature.
Summary of Contents for 2140-M Series
Page 1: ...WinCCS Lever Arm Tester Series 2140 M 2320 M 2330 M 2330 CC 2410 M 2410 CC...
Page 2: ...Revised March 2013...
Page 14: ...14...
Page 95: ...95 Once these steps are completed the thermocouples will be properly attached to the specimen...
Page 110: ...110...
Page 111: ...111...
Page 112: ...112...
Page 113: ...113...
Page 115: ...154 East Brook Lane Butler PA 16002 USA 1 724 283 1212 www atspa com...