76
4. With the lever arm in the level position, check balance and adjust it as necessary.
a. Loosen screws on each side of the lever arm counterweight.
b. Adjust the positioning screw on the counterweight for best lever arm balance, making it
slightly heavier on the weight pan side.
c. Tighten position locking screw and check balance.
5. Install lower load train and extensometer (if used).
6. Adjust lower take-up stud to ensure the lever arm is level when the load is applied.
7. Access the Start a Test screen:
a. From the main menu, select Tests, then Start. The Start a Test menu will appear.
b. From the main menu, select Specimens, then Start. The Start a Test menu will appear.
c. Click the Frame icon on the main menu.
d. Double-click on the Status Display line for the frame to be tested.
i. The Frame Services menu will show.
ii. Click on Start a Test, and the Start a Test menu will appear.
NOTE: In all four options, the correct frame must be selected by either double-clicking on it (Option a) or
selecting it from the Frame pull-down menu (Options b-d).
8. Verify that the Calibration/Verification Status section indicates that all required functions are
calibrated.
9. Verify that the Usage area of the Thermocouple Information section agrees with the thermocouple
installation for this particular test. Make changes if this does not agree.
Summary of Contents for 2140-M Series
Page 1: ...WinCCS Lever Arm Tester Series 2140 M 2320 M 2330 M 2330 CC 2410 M 2410 CC...
Page 2: ...Revised March 2013...
Page 14: ...14...
Page 95: ...95 Once these steps are completed the thermocouples will be properly attached to the specimen...
Page 110: ...110...
Page 111: ...111...
Page 112: ...112...
Page 113: ...113...
Page 115: ...154 East Brook Lane Butler PA 16002 USA 1 724 283 1212 www atspa com...