Technical Data
Incoming power:
Incoming power connection:
3P/N/PE
CEE (32A,5-pol.,6h)
Nominal voltage:
400V ±10%
Nominal frequency:
50/60 Hz
Rated curr:
max. 28 A
Nominal power:
max. 8 kVA
Nominal power of xenon lamps:
2200 VA
Power of the xenon lamps:
16.5 A
Source:
Light source:
2 xenon lamps, air-cooled
Lamp life (guaranteed):
1.500 h each
Sample chamber:
Capacity:
app. 2310 cm² (with XENOSENSIV)
Sample chamber temp.*:
20 °C – 65 °C
Black standard temperature*:
20 °C – 115 °C
Black panel temp.*:
20 °C – 110 °C
Relative humidity:
10% – 95%
Fan speed:
800 – 2500 rpm
Irradiance input limits:
Filter TM16
0.7 W/m² – 1.25 W/m² (420 nm)
Filter TM16
30 W/m² – 65 W/m² (300 – 400 nm)
XENOCHROME
®
320
30 W/m² – 90 W/m² (300 – 400 nm)
Daylight B04
30 W/m² – 120 W/m² (300 – 400 nm)
Water capacity:
Water tank capacity:
60l
Water quality:
Conductance:
<5μ S/cm
pH- value:
6 – 8
Solid content:
<1ppm
Silicate content
(measured as silicon dioxide SiO
2
):
<0.1ppm
Ultrapure water consumption:
Spraying: 0.7l/min
Humidity:
(100% activation oft the humidifier):
0.033 l/min
Connection conditions purified water:
Minimum pressure:
2 bar ±0.5 bar
Cooling air requirement:
Radiator cooling:
max. 150m
3
/h
Sample chamber:
max. 100m
3
/h
Interfaces:
RS232
Memory chip
Ethernet
Data recording:
RS232
Memory chip
Dimensions and weight:
Dimensions:
900 mm × 780 mm × 1800 mm
Empty weight (without operating media):
app. 290kg
Noise:
Noise level:
<70 db(A)
Ambient conditions:
Room temperature:
max. 18 °C – 25 °C
Relative humidity:
max. 50% (±10%)
Laboratory conditions:
Equipment with circulating air system
Volumetric flow of incoming air of at least 300m³/h
Air filtering with filter class EU4
dust-free
* The values depend on the lamp age, the test and
laboratory conditions.
11
Technical Data