Operating Manual Xenotest
®
440
- 12 -
Inhalt/Content/Sommaire
3.4
Water supply
The Xenotest
®
440 requires purified water for humidify
-
ing the test chamber. Purified water circulates in a closed
pipe system which is fed from the internal 60-liter water
tank. The water tank can either be refilled manually or
connected to a water treatment system.
• Connection: G 3/8 " (inch)
The purified water must have the following quality char
-
acteristics:
• Conductance: < 5 μS/cm
• pH value: 6 – 8
• Max. silicate content: 0.2 ppm
NOTE – water quality
Poor water quality not only leads to poor test results but
also affects the function and life of the components in
the instrument.
3
Requirements for the installation site
3.5
Transport
NOTE – Transport
The instrument may only be transported by experienced
service personnel using suitable equipment. The instru
-
ment may not be lifted by the front or side doors for
transport.
NOTE – lifting points (fig. 1)
The instrument may only be lifted at the lifting points
indicated in
fig. 1
.
Fig. 1