Operating Manual Xenotest
®
440
- 30 -
Inhalt/Content/Sommaire
6.3
Installing / removing the xenon lamps
NOTE – warranty complaint lamp set
In the event of a warranty claim, both lamps must be
sent back together with the appropriate warranty card
and the lamp file. Only then can a claim be made under
the warranty!
The lamp file of the used lamps can be saved on an SD
card under “Save settings/device/lamp age/lamp file.”
The file name corresponds to the serial number of lamp
set that is the subject of the complaint and must match
the warranty card and the lamp set. The file must be
saved before activating the new lamp set.
NOTE – XenoLogic™ function
The lamps must always be changed in pairs for the
XenoLogic function. Only then can a prolonged useful
lamp life be achieved.
WARNING – electrical shock!
Touching live parts can lead to life threat-
ening electrical shock. Disconnect the in-
strument from the power supply before
checking components!
• Turn the instrument off at the ON/OFF switch
• Pull the electrical plug out of the socket and secure
against reconnecting
• Check whether the instrument is voltage free
CAUTION – danger of burns!
The components of the lamp unit get very
hot and can cause burns when they come
into contact with the skin. Let the compo-
nents of the lamp unit cool down sufficient-
ly before changing the xenon lamps!
NOTE – handling the xenon lamps
Residue from finger marks can burn into the glass tubes
of the xenon lamps and cause uneven light radiation:
• Only hold the xenon lamps by the base
• Wear clean cotton gloves when handling the lamps
3
2
1
5
7
6
4
8
9
Fig. 19
6
Start up