Operating Manual Xenotest
®
440
Inhalt/Content/Sommaire
6.1
Initial start up
NOTE – first start up
The instrument may only be started for the first time by
an Atlas technical service or authorized service repre-
sentative. All the functioning components of the instru-
ment are checked during first-time start up.
6.2
Checking the instrument components
WARNING – electrical shock!
Touching live parts can lead to life threat-
ening electrical shock. Disconnect the in-
strument from the power supply before
checking components!
• Turn the instrument off at the ON/OFF switch
• Pull the electrical plug out of the socket and secure
against reconnecting
• Check whether the instrument is voltage free
Only instrument parts which are in optimal working condi
-
tion ensure the functional reliability of the Xenotest
®
440
instrument. Check the instrument components and op-
erational functions listed below for possible damage and
operability issues before starting or before every new test
run. Do not use damaged or worn components.
Checklist:
The following functions must be checked before starting:
• Correct combination of light filters
CAUTION – overheating!
The lamp filters assist in reducing heat ra-
diation. If tests are performed without fil-
ters, the instrument can overheat and the
samples will be damaged. Only ignite the
xenon lamps when filters are properly in-
stalled.
• Number of operating hours of the xenon lamps
• Condition of the air filters for lamp cooling and test
chamber cooling
• Fit and tightness of the water hoses
• Condition of the water filter
• Fill level of the instrument’s internal water tank and
supply tank for unpurified water
• Condition of plugs and power cables
6
Start up