- 13 -
Operating Manual Xenotest
®
440
Inhalt/Content/Sommaire
3
Requirements for the installation site
3.6
Space requirements
Position the instrument on a sufficiently stable, nonflam
-
mable floor and align horizontally.
Weight of the instrument:
• Xenotest
®
440
approx. 290 kg
NOTE – load at the installation site
If several instruments are to be installed in a room, en-
sure the static load-bearing capacity of the floor.
An access area of at least 1 m is needed around the
machine to allow for operation, maintenance and repair
work, see
fig. 2
(page 14).
• Instrument dimensions:
900 mm × 780 mm × 1800 mm (W × D × H)
CAUTION – overheating of the instrument!
The air exchange of the instrument takes
place via air inlet and outlet openings on
the top of the instrument and the side air
vents. If the air exchange is blocked, tests
will be interrupted because the safety de-
vices will shut the instrument off. Also, re-
peatedly turning the instrument on/off, will
shorten the life of instrument components.
Make sure that the ventilation openings are
always clear!
• The room of installation must have a minimum height
of 2.5 m
• Comply with the minimum distances at the side and
back of the instrument. The safety and maintenance
distances can be found in
fig. 2
(page 14)