![Atlas Xenotest 440 Operating Instructions Manual Download Page 28](http://html.mh-extra.com/html/atlas/xenotest-440/xenotest-440_operating-instructions-manual_3001605028.webp)
Operating Manual Xenotest
®
440
- 28 -
Inhalt/Content/Sommaire
5.9
Measuring and control sensors
Fig. 18
:
A combined temperature and humidity sensor
1
is in-
stalled in the test chamber of the Xenotest
®
440.
The test chamber temperature and relative humidity are
measured and regulated with this sensor.
The functions of the sensor are:
• Display of the measured test chamber temperature
• Constant control of the test chamber temperature ac
-
cording to programmed setpoint
• Display of the measured relative humidity in the test
chamber
• Constant control of the relative humidity in the test
chamber according to programmed setpoint
A XENOSENSIV
®
RC sensor for controlling the irradi
-
ance and surface temperature is positioned on the sam-
ple rack.
The functions of the XENOSENSIV RC sensor:
• Measurement of the surface temperature at sample
level
• Measurement of the irradiance at sample level
• Regulation of the surface temperature (BST or BPT)
according to a programmed setpoint
• Display of the measured irradiance
• Constant control of the irradiance according to pro
-
grammed setpoint
NOTE
The parameters mutually influence each other. A pre-
liminary test should be made to establish whether the
desired values are reached (see also section 7.4 and
section 7.5).
1
Fig. 18
5
Functional description