Operating instructions SunCool
TM
- 24 -
10
Specifications
10.1 SunCool
TM
units for SUNTEST CPS / CPS+ weathering units
Electrical system:
SunCool
TM
version 115V/60Hz:
• Mains voltage:
115 V ± 10%, 60 Hz
• Mains connection:
NEMA 5-15
• Starting current:
19 A (temporary)
• Current consumption
8 A (max.)
• Protection:
15 A
• Power consumption:
0.95 kW (max.)
SunCool
TM
version 230V/50Hz:
• Mains voltage:
230 V ± 10%, 50 Hz
• Mains connection:
(1, N, PE) AC or
(2, PE) AC
• Starting current:
19 A (temporary)
• Current consumption
5 A (max.)
• Protection:
16 A
• Power consumption:
0.95 kW (max.)
Refrigeration system:
SunCool
TM
version 115V/60Hz:
Refrigerant:
R 22
Capacity:
415 g
SunCool
TM
version 230V/50Hz:
Refrigerant:
R 407 C
Capacity:
450 g
Dimensions and weight:
• Dimensions in mm:
500 x 485 x 390
(Width x depth x height)
• Weight:
33 kg w/o connections
Noise emission:
• Noise level:
52 dB (A)
Performance:
SunCool
TM
version 115V/60Hz:
• Volume flow (room):
520 m
3
/h
• Cooling performance:
2.35 kW
SunCool
TM
version 230V/50Hz:
• Volume flow (room):
520 m
3
/h
• Cooling performance:
2.05 kW
Ambient conditions:
• Room temperature
22 °C to 40 °C.
• Relative humidity:
30% - 80%
• Air pressure:
700 - 1060 hPa