- 17 -
Operating instructions SunCool
TM
Figure 13
5.2
Settings
Depending on the unit version and control panel used, the
following settings are required for an operation with the
SUNTEST weathering unit. Please note that either the con-
nection line or the dummy plug must be installed when the
SunCool
TM
is operated with a SUNTEST (see section 5.3).
5.2.1 Version 115V / 60Hz
NOTE - Test results:
Accurate test results can only be obtained with the set-
tings specified by ATLAS. The SunCool
TM
has a settling
time of approximately ten minutes before reaching maxi-
mum cooling performance.
To turn the SunCool
TM
on:
Rotate the
THERMOSTAT
switch to position
9
.
Rotate the
OPERATION
switch to position
COOL
onto
the
large snowflake
symbol.
To turn the SunCool
TM
off:
•
When the SunCool
TM
is operated with the SUNTEST
CPS+ / XLS+, it is turned off by the program control.
However, for extended operation interruptions, the rota-
ry switch OPERATION should be in position 0.
•
For operation with the SUNTEST CPS / XLS without
program control, turn the rotary switch OPERATION to
position 0 to turn the SunCool
TM
off.
5.2.2 Version 230V / 50Hz
NOTE - Test results:
Accurate test results can only be obtained with the set-
tings specified by ATLAS.
To turn the SunCool
TM
on:
Rotate the
THERMOSTAT
switch to position
11
.
Rotate the
FAN SPEED
switch to position
HIGH
(high fan
speed).
Set the
OPERATION
switch to position ON.
To turn the SunCool
TM
off:
•
When the SunCool
TM
is operated with the SUNTEST
CPS+ / XLS+, it is turned off by the program control.
However, for extended operation interruptions, the switch
OPERATION should be in position OFF.
•
For operation with the SUNTEST CPS / XLS without
program control, set the switch OPERATION to positi-
on OFF to turn the SunCool
TM
off.
5
Operation and shut-down
CAUTION - Danage to the cooling system:
To prevent the cooling system from being damaged,
the SunCool
TM
must remain turned off for at least
three minutes before it is turned on again.
1
Rotary switch THERMOSTAT in position 9
2
Rotary switch OPERATION in position COOL
(symbol with large snowflake)
OPERATION
THERMOSTAT
0
F
AN
COOL
1
9
8
7
6
5
4
3
2
C
O
O
L
E
R
BEFORE RESTARTING
WAIT THREE MINUTES
SunCool 115 V / 60Hz
1
2
1
Rotary switch THERMOSTAT in position 11
2
Switch FAN SPEED in position HIGH
3
Switch OPERATION in position ON
OPERATION
OFF
ON
THERMOSTAT
WAIT 3 MINUTES BEFORE RESTARTING
3
1
5
7
9
11
COOLER
FAN SPEED
LOW
HIGH
SunCool 230 V / 50Hz
3
1
2