Reference
3-18
GB1400 User Manual
Procedure to Control Error Injection Mode
1.
Press the RATE key one or more times to select the desired error injection
mode. Note that the LED in the RATE key will turn on except when mode is
set to ERR OFF. Available selections are:
ERR OFF: Single error injection.
ERR 1E-7: Generate errors at rate of 10-7.
ERR 1E-6: Generate errors at rate of 10-6.
ERR 1E-5: Generate errors at rate of 10-5.
ERR 1E-4: Generate errors at rate of 10-4.
ERR 1E-3: Generate errors at rate of 10-3.
ERR EXT: External error inject mode.
2.
Once you have selected an internal error rate, or the external mode, you can
turn this error rate off and on by alternately pressing the SINGLE key and
RATE key. Do not press the RATE key two or more times in a row unless
you want to change the current error injection mode. However, while the
current error rate is off, you can press the SINGLE key as many times as you
wish to inject single errors.
3.
To return to the single injection mode, press RATE one or more times until
ERR OFF is selected.
ERROR INJECT Input
The ERROR INJECT input is an SMA (female) connector located on the rear-
panel of the Generator. When the error injection mode is set to ERR EXT, one bit
error will be generated for each rising edge in the signal applied to this input.
Setup
- The ERROR INJECT is a 50 Ohm, ECL input. No hardware setup is
required.
Data Inhibit
- The DATA INHIBIT input is located on the rear-panel of the
Analyzer. A signal applied to this input may be used to gate off and on the
Generator DATA output signal. The logic of the DATA INHIBIT function is
shown in the following table. The DATA INHIBIT function is not bit or frame
synchronized with the DATA output signal. Therefore, the gating action caused
by the DATA INHIBIT input may occur anywhere within a DATA output bit
time slot and anywhere within a pattern frame. DATA INHIBIT is a standard 50
Ohm ECL input and does not require any threshold or delay setup.
Table 3-3. Data Inhibit Logic
Logic Level Applied to
DATA INHIBIT INPUT
Action
OPEN or LOW
DATA output operates normally.
HIGH
DATA output is disabled, that is forced LOW.
Summary of Contents for gigaBERT1400
Page 16: ...Preface GB1400 User Manual xvi ...
Page 17: ...Getting Started ...
Page 26: ...Operating Basics ...
Page 66: ...Application Note Auto Search Synchronization 2 40 GB1400 User Manual ...
Page 76: ...Application Example 2 50 GB1400 User Manual ...
Page 77: ...Reference ...
Page 163: ...Reference 3 86 GB1400 User Manual ...
Page 164: ...Appendices ...
Page 174: ...Specifications A 10 GB1400 User Manual ...
Page 194: ...BERT Technical Articles B 20 GB1400 User Manual ...
Page 320: ...Remote Commands C 102 GB1400 User Manual ...
Page 332: ...Using GPIB RS 232 D 12 GB1400 User Manual ...
Page 341: ...Performance Verification GB1400 Acceptance Test E 9 ...
Page 349: ...Default Settings F 6 GB1400 User Manual MISC View Angle 0 Panel Lock OFF Response Header ON ...
Page 351: ...Cleaning Instructions G 2 GB1400 User Manual ...
Page 368: ...Glossary Index ...