Functional Overview
GB1400 User Manual
2-7
Analyzer (RX) Display
Like the Generator, the Analyzer has a two-line by 24-character high-contrast
liquid crystal display (LCD).The Analyzer display in its normal (non-menu)
mode is illustrated in the figure below.
Frequency (kHz)
Error Rate
Totalize
1405000
5.0E-06
2410538
1.2 ns
PN23
-0.05 V
Delay/ Memory
Figure 2-8. Analyzer Display in Its Normal (Non-menu) Mode
Like the Generator, the Analyzer has a two-line by 24-character high-contrast
liquid crystal display (LCD). The function of each field in the normal Analyzer
display format -that is the format used when the Analyzer is not in the menu
mode -is described below:
•
The
top left
section of the Analyzer display is used to show the measured
frequency of the input clock signal in MHz. For example a display of
6 2 2 . 0 5
indicates a measured frequency of 622.05 MHz. Note that the
Analyzer frequency display contains five significant digits while the
Generator frequency display contains six. This is because the frequency
shown in the Analyzer display is a
measurement
result
while the frequency
shown in the Generator display is an
instrument
setup
which is known with
more precision.
•
The
top
middle
and
top right
sections of the display normally show measured
bit error rate and bit errors respectively. BER is expressed in exponential or
"E" notation. For example, a display of 1.5E-09 indicates a measured BER
of 1.5 x 10-9. The Analyzer calculates BER and counts bit errors in three
modes simultaneously:
Window, Test, and Totalize
. The symbol in front of
the BER field indicates which mode has been selected for display.
Window
results are preceded by a blank space, that is no symbol.
Totalize
results will
be preceded by an
∞
(infinity) symbol.
Test results
will be preceded by a
T
,
U
, or
R
depending on the selected test mode: timed, untimed, or repeat.
Refer to Chapter 4 for more information on displaying Analyzer results and
starting and stopping tests.
Summary of Contents for gigaBERT1400
Page 16: ...Preface GB1400 User Manual xvi ...
Page 17: ...Getting Started ...
Page 26: ...Operating Basics ...
Page 66: ...Application Note Auto Search Synchronization 2 40 GB1400 User Manual ...
Page 76: ...Application Example 2 50 GB1400 User Manual ...
Page 77: ...Reference ...
Page 163: ...Reference 3 86 GB1400 User Manual ...
Page 164: ...Appendices ...
Page 174: ...Specifications A 10 GB1400 User Manual ...
Page 194: ...BERT Technical Articles B 20 GB1400 User Manual ...
Page 320: ...Remote Commands C 102 GB1400 User Manual ...
Page 332: ...Using GPIB RS 232 D 12 GB1400 User Manual ...
Page 341: ...Performance Verification GB1400 Acceptance Test E 9 ...
Page 349: ...Default Settings F 6 GB1400 User Manual MISC View Angle 0 Panel Lock OFF Response Header ON ...
Page 351: ...Cleaning Instructions G 2 GB1400 User Manual ...
Page 368: ...Glossary Index ...