Reference
3-58
GB1400 User Manual
Word Fill (FILL)
Format:
F1:ESC F4:SET
FILL WORD MEMORY WITH:hh
Menu:
WORD
Function Name:
FILL
Instruments:
Analyzer and Generator
Options:
Requires the 1-Mbit Option.
Application:
Use this function to fill all bytes in the current WORD with the
same 8-bit pattern.
Parameters:
Fill Byte (hh): Enter the hex value for the fill byte Range: 00 to
FF.
Notes:
You may use the fill function as the basis for a new WORD
pattern, then edit individual bytes using the WORD EDIT
function to create the exact pattern that you need.
See an additional list of remote commands in the Appendix that
support the 1-Mbit Programmable Word option.