Performance Verification
GB1400 Acceptance Test
E-3
Functional Test
To functionally test the GB1400, connect the Generator to the Analyzer and confirm
their correct operation as described below.
Note
In these procedures, the Generator and Analyzer are returned to their default
settings. The word memories will be reset. If you have entered word patterns
that you do not want to lose, use a GPIB or RS-232C controller to save them
before beginning these procedures
.
q
Step 1:
Make connections appropriate to the options installed as follows:
Connections for Functional Test
Option
From
To
Comment
Standard Instrument
TX Clock
RX Clock
50 Ohm, SMA coax
TX Not Clock
RX Not Clock
50 Ohm, SMA coax
TX Data
RX Data
50 Ohm, SMA coax
TX Not Data
RX Not Data
50 Ohm, SMA coax
75 Ohm Option
TX Clock
RX Clock
TX Not Clock
RX Not Clock
TX Data
RX Data
TX Not Data
RX Not Data
50 Ohm, SMA coax.
Use SMA to BNC
adapters and 75 Ohm
BNC coaxial cables
See Figure 1 for Setup of Functional Test.
q
Step 2:
Reset both the Generator and Analyzer to their factory default settings.
Press the ‘CLEAR’ key in the PATTERN section, the ‘MSB(1)’ key and the ‘VIEW
ANGLE’ key, and hold down all three keys while cycling the power.
The message DEFAULT SETTINGS should appear briefly on the display of each unit,
followed by the GB1400 logo, before the functional display appears.
q
Step 3:
At the Generator, verify that error injection is off (you should see ERR OFF
on the display). If error injection is on, press the ‘SINGLE’ key in the ERROR
INJECT section to turn it off.
Summary of Contents for gigaBERT1400
Page 16: ...Preface GB1400 User Manual xvi ...
Page 17: ...Getting Started ...
Page 26: ...Operating Basics ...
Page 66: ...Application Note Auto Search Synchronization 2 40 GB1400 User Manual ...
Page 76: ...Application Example 2 50 GB1400 User Manual ...
Page 77: ...Reference ...
Page 163: ...Reference 3 86 GB1400 User Manual ...
Page 164: ...Appendices ...
Page 174: ...Specifications A 10 GB1400 User Manual ...
Page 194: ...BERT Technical Articles B 20 GB1400 User Manual ...
Page 320: ...Remote Commands C 102 GB1400 User Manual ...
Page 332: ...Using GPIB RS 232 D 12 GB1400 User Manual ...
Page 341: ...Performance Verification GB1400 Acceptance Test E 9 ...
Page 349: ...Default Settings F 6 GB1400 User Manual MISC View Angle 0 Panel Lock OFF Response Header ON ...
Page 351: ...Cleaning Instructions G 2 GB1400 User Manual ...
Page 368: ...Glossary Index ...