Tutorial
GB1400 User Manual
2-33
B.
Locate the controls in the
PATTERN
box of the Generator. Make sure the
PRBS button is enabled (the
LED
inside this switch should be
ON
). Set the
Generator pattern to a
223-1
bit
PRBS
as follows:
Set this parameter
…to this value
…using this procedure.
PRBS type
PN 23
Press pattern up/down arrow keys until
PATTERN is set to PN 23
C.
Locate the controls in the
ERROR INJECT
box of the Generator. Verify
that the Generator
ERROR RATE GENERATOR
is
OFF
(
the
LED
within the switch should
be
OFF
). If the LED in the error inject RATE key is ON, then press RATE one or
more times until it turns off.
4.
Setup Analyzer for “AUTO-SEARCH” Operation.
A.
Locate the controls in the
SYNC
box of the Analyzer. Verify that the
Analyzer
AUTO SEARCH
function is
ENABLED.
The
LED
in the
AUTO SEARCH
key
should be
ON
. If the LED is OFF, press the
AUTO SEARCH
function one time until
the
LED
is
O N
. At this point, verify that the green
LOCK LED
is
ON
.
B.
Locate the controls in the
ERROR DETECTION
box of the Analyzer. Zero
all Analyzer error counts by pressing the
CLEAR
key. Confirm that the number of
errors and the error rate were reset to 0 (note: error rate will start changing as
more and more bits are received. After several minutes of operation, the error
rate should reach 0.0E-9
⇒
0.0E-10
⇒
0.0E-11 and on).
C.
Locate the controls in the
ERROR HISTORY
box of the Analyzer. Reset all
Analyzer history LEDs by pressing this
CLEAR
key. Confirm that all
ERROR
HISTORY LED’s
are turned OFF.
D.
Locate the controls in the
ERROR INJECT
box of the Generator. Verify the
GB1400 Analyzer can detect errors by pressing the Generator
ERROR INJECT
SINGLE
key several times. Verify that the number of errors count displayed by
the Analyzer increments each time the Generator
SINGLE
key is pressed.
In effect you are now performing a bit error rate test on the test cables connecting
the
gigaBERT
1400 Generator and Analyzer. In an actual BER test,
gigaBERT
1400 Generator clock and data outputs would be connected to inputs
on a "device under test" (DUT) while
gigaBERT
1400 Analyzer inputs would be
connected to outputs on the DUT.
Summary of Contents for gigaBERT1400
Page 16: ...Preface GB1400 User Manual xvi ...
Page 17: ...Getting Started ...
Page 26: ...Operating Basics ...
Page 66: ...Application Note Auto Search Synchronization 2 40 GB1400 User Manual ...
Page 76: ...Application Example 2 50 GB1400 User Manual ...
Page 77: ...Reference ...
Page 163: ...Reference 3 86 GB1400 User Manual ...
Page 164: ...Appendices ...
Page 174: ...Specifications A 10 GB1400 User Manual ...
Page 194: ...BERT Technical Articles B 20 GB1400 User Manual ...
Page 320: ...Remote Commands C 102 GB1400 User Manual ...
Page 332: ...Using GPIB RS 232 D 12 GB1400 User Manual ...
Page 341: ...Performance Verification GB1400 Acceptance Test E 9 ...
Page 349: ...Default Settings F 6 GB1400 User Manual MISC View Angle 0 Panel Lock OFF Response Header ON ...
Page 351: ...Cleaning Instructions G 2 GB1400 User Manual ...
Page 368: ...Glossary Index ...