Reference
3-26
GB1400 User Manual
Input Decision Threshold
The input decision thresholds of the DATA, and REF DATA inputs can be
independently adjusted. However, it is important to note that input threshold
adjustment applies only for
single-ended operation
. When operating Analyzer
data inputs in the differential mode, their input decision threshold effectively
becomes the average of the positive peak and negative peak voltage levels and is
not adjustable by the user.
Procedure to Adjust (Single-ended) Input Decision Threshold:
1.
Verify that the selected input is set up for singled-ended operation. For
DATA operation, connect a cable from rear panel threshold to front NOT-
DATA (DATA-BAR) input.
2.
Select either the DATA, CLOCK, or REF DATA input using the F2 and F3
keys.
3.
Press the THRESH key. Verify that its LED turns on.
4.
Press the input up/down keys while observing selected threshold value in the
bottom, right field of the display. Decision threshold setup range is a function
of input termination as shown in table below.
Table 3-8. Input Threshold Range as a Function of Termination
Selected Termination
Threshold Setup Range
GND
- 1.5 to + 1.0 VDC, 50 mV steps
- 2 V
- 2.5 to + 0.0 VDC, 50 mV steps
AC
- 1.5 to + 1.0 VDC, 50 mV steps
The displayed threshold takes effect immediately. In other WORDs, threshold is
incremented or decremented each time you press the up or down key. Note that
threshold is an AUTO SEARCH parameter. So be sure turn off AUTO SEARCH
if you want to fix delay at a specific value.
Logically Inverting Input Data
The INPUT section INVERT key may be used to logically invert the input data
pattern, that is change all 1s to 0s and 0s to 1s. If the INVERT key is off, then the
input data pattern is not inverted. However if the INVERT key is on, the input
data pattern is logically inverted. You can toggle input data inversion on or off by
pressing the INVERT DATA key. Note that logically inverting the input pattern
is equivalent to swapping the connections to the DATA and DATA-BAR inputs.
To toggle input data inversion on or off, press the INVERT key.
Summary of Contents for gigaBERT1400
Page 16: ...Preface GB1400 User Manual xvi ...
Page 17: ...Getting Started ...
Page 26: ...Operating Basics ...
Page 66: ...Application Note Auto Search Synchronization 2 40 GB1400 User Manual ...
Page 76: ...Application Example 2 50 GB1400 User Manual ...
Page 77: ...Reference ...
Page 163: ...Reference 3 86 GB1400 User Manual ...
Page 164: ...Appendices ...
Page 174: ...Specifications A 10 GB1400 User Manual ...
Page 194: ...BERT Technical Articles B 20 GB1400 User Manual ...
Page 320: ...Remote Commands C 102 GB1400 User Manual ...
Page 332: ...Using GPIB RS 232 D 12 GB1400 User Manual ...
Page 341: ...Performance Verification GB1400 Acceptance Test E 9 ...
Page 349: ...Default Settings F 6 GB1400 User Manual MISC View Angle 0 Panel Lock OFF Response Header ON ...
Page 351: ...Cleaning Instructions G 2 GB1400 User Manual ...
Page 368: ...Glossary Index ...