![ATAGO 3275 Instruction Manual Download Page 71](http://html.mh-extra.com/html/atago/3275/3275_instruction-manual_2995787071.webp)
69
Instrument Specifications
(Common Specifications on Page 70 "Common specifications")
Automatic Digital Refractometer RX-5000i-Plus Cat.No.3275
Measurement range
① Refractive Index (nD): 1.32420 to 1.58000 ② Brix : 0.000 to 100.000%
Resolution
① Refractive Index (nD): 0.00001 ② Brix: 0.001%
③ Temperature: 0.01℃
Measurement accuracy ① Refractive Index (nD):
±
0.00002 ② Brix:
±
0.010%
(When measuring a standard sucrose solution of up to 50% Brix or standard
refractive index solution in MODE-1or MODE-T at 20℃)
Repeatability
① Refractive Index (nD):
±
0.00001 ② Brix:
±
0.010%
(When measuring a standard sucrose solution of up to 50% Brix or standard
refractive index solution in MODE-1or MODE-T at 20℃)
Dimensions and weight 37
×
26
×
14cm, 6.6kg (Main Unit only)
Automatic Digital Refractometer RX-5000i
Cat.No.3276
Measurement range
① Refractive Index (nD): 1.32420 to 1.58000 ② Brix : 0.00 to 100.00%
Resolution
① Refractive Index (nD): 0.00001 ② Brix: 0.01% ③ Temperature: 0.01℃
Measurement accuracy ① Refractive Index (nD):
±
0.00004 ② Brix:
±
0.03%
(When measuring a standard sucrose solution of up to 50% Brix or standard
refractive index solution in MODE-1 at 20℃)
Repeatability
① Refractive Index (nD):
±
0.00002 ② Brix:
±
0.01%
(When measuring a standard sucrose solution of up to 50% Brix or standard
refractive index solution in MODE-1 at 20℃)
Dimensions and weight 37
×
26
×
14cm, 6.6kg (Main Unit only)
Automatic Digital Refractometer RX-7000i
Cat.No.3279
Measurement range
① Refractive Index (nD): 1.29980 to 1.71500 ② Brix : 0.00 to 100.00%
Resolution
① Refractive Index (nD): 0.00001 (Factory default setting = 0.0001)
② Brix: 0.01% (Factory default setting =0.1%) ③ Temperature: 0.01℃
Measurement accuracy ① Refractive Index (nD):
±
0.0001 ② Brix:
±
0.1% *1
Repeatability
① Refractive Index (nD):
±
0.00005 ② Brix:
±
0.02% *1
Dimensions and weight 37
×
26
×
14cm, 7.0kg (Main Unit only)
*1 When measuring a standard sucrose solution of up to 50% Brix or standard refractive index solution in
MODE-1 at 20℃. Guaranteed accuracy range is nD 1.32420 to 1.70000 and Temperature 5.0 to 70.0
°
C.
Automatic Digital Refractometer RX-9000i
Cat.No.3278
Measurement range
① Refractive Index (nD): 1.29980 to 1.71500 ② Brix : 0.00 to 100.00%
Resolution
① Refractive Index (nD): 0.00001 ② Brix: 0.01% ③ Temperature: 0.01℃
Measurement accuracy ① Refractive Index (nD):
±
0.00004
*2
② Brix:
±
0.03%
*3
③ Brix:
±
0.05%
*4
*1
Repeatability
① Refractive Index (nD):
±
0.00002
*2
② Brix:
±
0.01%
*3
③ Brix:
±
0.01%
*4
*1
Dimensions and weight 37
×
26
×
14cm, 7.0kg (Main Unit only)
*1 When measuring a standard sucrose solution of up to 50% Brix or standard refractive index solution in
MODE-1. Guaranteed accuracy range is nD 1.32420 to 1.70000 and Temperature 5.0 to 70.0
°
C.
*2 When nD 1.33299 to 1.42009, 10.00 to 30.00℃. For other range, nD ±0.00010 (Repeatability nD±
0.00005).
*3 When Brix 0.00 to 50.00%, 10.00 to 30.00℃.
*4 When Brix 50.01 to 95.00%, 10.00 to 30.00℃. For other range, Brix ±0.10%(Repeatability Brix ±
0.02%).