64
6a.
The instrument is
functioning properly if
GOOD is displayed. Touch
"ENTER" to return to the
home screen.
6b.
If ERROR is displayed,
clean the prism thoroughly
and touch "ENTER" to
perform the test again.
6c.
If ERROR continues to
appear, contact ATAGO.
※
When ERROR is displayed,
perform the following
operations in MODE-1 at
20℃:
①
Check if purified water
measures within the
accuracy range of
±
0.010%
(RX-5000i-Plus) or
±
0.03%
(RX-5000i/RX-9000i) or
±
0.1% (RX-7000i).
②
Zero-set and confirm the
zero-setting (see page 26-
"Zero-setting").
③
Check if standard sucrose
solutions measure within the
accuracy range.
※
"Keep the prism clean" is
displayed.
※
"Clean the prism and press
'ENTER'" is displayed.