![ATAGO 3275 Instruction Manual Download Page 68](http://html.mh-extra.com/html/atago/3275/3275_instruction-manual_2995787068.webp)
66
Cleaning Fan and Replacing Dust Filter
Dusty cooling fans, the intake fan on the back and the exhaust fan at the bottom, may cause malfunction of
the instrument. Clean the fans regularly and replace the filter as necessary.
(1)
Intake cooling fan (back)
is located on the backside of the
instrument.
Remove the cover with a [type of driver
need to be identified] screw driver.
Clean the filter under warm soapy water
or replace with a new filter.
Place the cover back on and secure.
(2)
Exhaust cooling fan (bottom)
Never remove the cover. Put the nozzle of
a vacuum cleaner against it to clean the
dust out.
Filter replacement alert
After a period of usage, the instrument will automatically alert the user to replace the dust filter. Follow the
below procedure after cleaning/replacing the dust filter.
1.
Touch the dust filter
indication to open the system
information screen.
2.
Touch "CLEAR" to reset the
usage time tracker.
※
Uncheck "Check Fan filter"
to turn off the filter
replacement alert feature.
Intake cooling fan
Hook
Filter
Filter cover
Exhaust cooling fan
Summary of Contents for 3275