4.
Select either the Brix or
refractive index scale.
5.
"SET SAMPLE AND PRESS"
will be displayed.
6.
Place the standard solution,
close the cover plate, and
touch "ENTER.
”
7.
Measurement is taken once
the target temperature is
reached.
8.
Once the measurement value
is displayed, adjust SETTING,
using the arrows ▲▼.
Touch "ENTER" to save the
settings.
※
Make sure that the instrument is set to MODE-1; otherwise, "MANUAL
CALIBRATION" will not launch in the step 3.
※
In case of a measurement error, OUT OF CALIBRATION will be displayed.
※
Once the SETTING number is changed from "000," the steps 4 through 7 will be
skipped from next time. The screen in the step 8 will be displayed immediately when
"MANUAL CALIBRATION" is selected in the step 3.
※
In order to change the scale to adjust, change the SETTING number to "000" in the
step 8 and touch "ENTER.
”
The next time "MANUAL CALIBRATION" is launched,
the scale selection screen will appear.
※
While the Peltier is in operation, a
WAITING message will be
displayed. Touch "QUIT" to abort
the operation.