![ATAGO 3275 Instruction Manual Download Page 64](http://html.mh-extra.com/html/atago/3275/3275_instruction-manual_2995787064.webp)
62
Error message
Cause
Action
The instrument did not power on
correctly.
Turn the power off and wait for at least 1
minute to turn it back on. If the error
message continues to appear, turn the power
off, unplug the power cable, and contact
ATAGO.
The thermo module did not start
correctly when powered on.
Turn the power off and wait for at least 1
minute to turn it back on. If the error
message continues to appear, turn the power
off, unplug the power cable, and contact
ATAGO.
The thermo module did not start
correctly during mode setting.
Re-do the mode setting. If the error message
continues to appear, turn the power off,
unplug the power cable, and contact ATAGO.
A wrong password was entered;
or an operation was attempted
by an unauthorized user (in
system levels 2 & 3).
Enter the correct password only for
authorized operations.
Error in saving data on USB
flash drive
See page 41 "Saving Data on USB Flash
Drive"
The screen suddenly turns dark.
⇒
The LCD dims automatically after 5 minutes of non-operation. There is no setting to disable this
feature. Simply touch the screen to restore the brightness.
The administrator's password has been lost.
⇒
Contact ATAGO.
Summary of Contents for 3275
Page 2: ......
Page 74: ...1811K Printed in Japan ...