Warning of safety defects due to improper
connections of the ME system
Connection of the ME system with other de-
vices or installations or pieces of equipment
not recommended by ASSKEA GmbH and
not specified in the instruction for use may
compromise the safety and function of the
ME system. Any warranty is excluded for
damage caused by connecting devices or
installations or pieces of equipment not re-
commended with the ASSKEA secretion as-
pirator M30
Plus
or by improper use.
Connect only recommended original parts
with the ASSKEA secretion aspirator M30
Plus
.
Inspection of the internal power supply
Since the internal battery is not automatically
kept in a fully operational condition, the state
of charge must be checked periodically and,
if necessary, the battery must be replaced by
service staff. The battery may be replaced
only by authorized service staff, since re-
placement by inadequately trained persons
could result in a hazard (such as excessive
temperatures, fire or explosion)!
Damage to the device from ingress of liquids
The ASSKEA secretion aspirator M30
Plus
has
the IP classification IP22 against the ingress
of liquids. Nevertheless protect the device
from moisture (see symbol in section 1.2.1).
• Do not use the device near splashing water.
• Do not use the device in damp rooms or
while bathing or showering.
• Do not allow the power supply unit, push-
button or the socket for the power supply
unit to get wet.
• Never submerge the device in water or
other liquids (even when it is switched off).
Known or identifiable conditions when using
the device for medical care within a home
environment
• Children and pets must be kept away from
the device. Ensure that the device cannot
be knocked over or fall in such a way that
persons can be hit by the falling device.
• Prior to connecting the power supply unit,
ensure that the voltage of the device corre-
sponds to the home’s power supply.
• Do not use the device in damp rooms, ba-
throoms or showers.
• Do not allow the power supply unit, push-
button and the socket for the power supply
unit to get wet.
• Never submerge the device in water or
other liquids (even when it is switched off).
• Ensure that lint and dust are immediately
removed from the device and accessories
in order to ensure unimpaired functionality.
In addition, vermin must not be allowed in
the proximity of the device, as there is a risk
they may get inside the device and cause
damage.
• Never expose the device or accessories
to direct sunlight as this could result in a
strong heating effect, causing functional
impairments.
• Some devices and sources that are com-
monly used in the home environment may
cause potential faults in the device or ac-
cessories, e.g. fireplaces or radiant heaters
(strong heating of the device) or inhalers or
steam kettles (excessive air humidity). Do
not operate such devices and sources in the
vicinity of the ASSKEA secretion aspirator
M30
Plus
.
Possible physiological effects and unobvious
risks
• Select the vacuum depending on the res-
pective patient and on the medical indica-
tion to avoid personal injuries.
• Never use the ASSKEA secretion aspirator
M30
Plus
to aspirate solids or inhaled foreign
objects out of the throat. This could lead to
the object being pushed deeper into the air-
ways potentially causing total obstruction.
• Always place the device upright on a sturdy
and flat base without slope. Make sure that
the device cannot be knocked over or fall in
such a way that persons could be hit by the
falling device.