31
ENGLISH
5 Troubleshooting
Error
LED display
Possible causes
Remedy
Device does
not start.
LEDs do not
illuminate
Battery is discharged.
Connect the power supply unit for charging
or operating.
LED blinks red
Battery is discharged and invalid input
voltage
Connect the correct power supply unit;
check the power supply.
Device does
not charge.
LED blinks red
Temperature of the battery too high for
charging
No measure required, automatic resumption of
charging when battery has cooled down.
LED illuminates
green
Battery is defective.
Please contact Service!
LED blinks red
Invalid input voltage
Connect the correct power supply unit,
check the power supply.
Diminished
flow rate
LED blinks
green/yellow
Device is overheated and is only running in
curbed operation.
Switch off the device and allow it to cool down.
Insufficient
flow rate
LED illuminates
green/yellow
External bacterial filter blocked
Replace the external bacterial filter.
Integrated bacterial filter blocked
Please contact Service!
LED illuminates
green
Leak in the tubing system or in the canister
system
Check that secretion canister lid and tubing connections
are tight and no leakages are apparent.
LED illuminates
red
Battery is weak.
Connect the power supply unit for charging
or operating.
LED does not
illuminate