AZI Customer Service 800-528-7411 Manual, ZSP-150-F Page 10 of 22
The RANGE/FUNCTION Button
Pressing the RANGE/FUNCTION button begins the following menu cycle:
•
PCS – Front panel parts counting
See Appendix A (p. 18) for details
•
HI OK LO – Checkweighing
See Appendix B (p. 19) for details
•
CAL 1 – Auto-calibration
See Auto-calibration on p. 10
•
CAL 2 – Linearity – Factory Use Only
DO NOT USE
•
T –
Live animal weighing
See Appendix C (p. 20) for details
•
% - Percent weighing
See Appendix D (p. 21) for details
To select the desired function, press the MODE button when it appears on the display.
CAL 2 is set at the factory and should never require further adjustment. The procedure for CAL 1
begins below. For the other functions listed in the RANGE/FUNCTION menu cycle, refer to the
appendices, beginning on page 18.
CAL 1 [Auto-Calibration] - Using an external calibration weight
Perform auto-calibration every time you move your balance. It is extremely important to use high
quality weights and verify the balance has been warmed up for at least one (1) hour prior to
calibration.
User Action
Balance Response
1 Remove any containers or weighing
samples so that nothing is on the weighing
pan, then press the ZERO button.
Zeros are displayed.
2 Press the RANGE/FUNCTION button
Balance display cycles repeatedly through PCS, HI
OK LO, CAL 1, CAL 2,
T
and %.
3 Press the MODE button when CAL 1
appears.
CAL 1 and a flashing 0 are displayed.
4 Wait 10 seconds for the balance to
stabilize, then press the ZERO button.
The display stops flashing and a solid 0 is displayed.
In approximately 15 seconds, when this step is
complete, a single weight or two alternating weights
will begin flashing on the display.
5 Place one of the flashing weights in the
center of the balance’s weighing pan.
The display stops flashing the alternating weights
and the selected weight is displayed.
6 Wait 10 seconds for the balance to
stabilize, then press the ZERO button.
The display will momentarily blink, then the selected
weight will continue to be displayed. In approximately
15 seconds, when this step is complete, the display
will blank, flash “OK”, then display the calibration
weight including decimal places. The balance is now
calibrated and in the normal weighing mode.
7 Remove the calibration weight.
The display returns to zeros.