AZI Customer Service 800-528-7411 Manual, ZSP-150-F Page 19 of 22
Appendix B - Weight Checking
The following procedure checks the weight of individual parts against user-programmed upper and
lower weight limits.
Procedure
Display Reading
1 Press the RANGE/FUNCTION
button.
Display cycles repeatedly through PCS, HI-OK-LO,
CAL1, CAL2,
T
and %.
2 Press the MODE button when HI-OK-
LO appears on the display.
HI, LO and O flash.
3 Press the ZERO button.
HI flashes and zeros are displayed.
4 Place weight on pan that represents
the upper limit of the acceptable
weight span.
HI flashes and weight is displayed.
5 Press the ZERO button.
LO flashes and upper limit weight is displayed.
6 Remove upper limit weight and place
weight on the pan that represents the
lower limit.
LO flashes and weight is displayed.
7 Press the ZERO button.
HI-OK-LO briefly appears then OK and the low limit
weight are displayed.
8 Remove low limit weight.
LO appears and zeros are displayed.
The balance is now ready to check the weight of parts. The balance displays OK along with the
weight of the part if the part is within the selected limits. If the part is too heavy, HI and the weight are
displayed. If the part is too light, LO and the weight are displayed. The limits are stored until the
balance is turned off. To exit weight checking, press the MODE button and the balance will return to
normal operation.